Successful Feasibility Human Trial of a New Self-Expandable Percutaneous Pulmonary Valve (Pulsta Valve) Implantation Using Knitted Nitinol Wire Backbone and Trileaflet-Gal-Free Porcine Pericardial Valve in the Native Right Ventricular Outflow Tract

Gi Beom Kim, Mi Kyoung Song, Eunjung Bae, Eun Ah Park, Whal Lee, Hong Gook Lim, Yong Jin Kim

Research output: Contribution to journalArticleResearchpeer-review

3 Citations (Scopus)

Abstract

Background-Self-expandable percutaneous pulmonary valve implantation (PPVI) for native right ventricular outflow tract lesions is still in the clinical trial phase. The aim of this study is to present the result of feasibility study of a novel self-expandable knitted nitinol wire stent mounted with a treated trileaflet-Gal-free porcine pericardial valve for PPVI. Methods and Results-A feasibility study using Pulsta valve (TaeWoong Medical Co, Gyeonggi-do, South Korea) was designed for patients with severe pulmonary regurgitation in the native right ventricular outflow tract, and 6-month follow-up outcomes were reviewed. Ten tetralogy of Fallot patients were enrolled. Before PPVI, severe pulmonary regurgitation (mean pulmonary regurgitation fraction, 45.5%±7.2%; range, 34.9%-56%) and enlarged right ventricular volume (mean indexed right ventricular end-diastolic volume, 176.7±14.3 mL/m2; range, 158.9-205.9 mL/m2) were present. The median age at PPVI was 21.7±6.5 years (range, 13-36 years). Five patients were successfully implanted with 28 mm and the other 5 with 26 mm valves loaded on the 18F delivery cable. No significant periprocedural complications were noted in any patient. At the 6-month follow-up, indexed right ventricular end-diastolic volume was dramatically decreased to 126.3±20.3 mL/m2 (range, 99-164.2 mL/m2), and the mean value of peak instantaneous pressure gradient between the right ventricle and the pulmonary artery decreased from 6.8±3.5 mm Hg (range, 2-12 mm Hg) before PPVI to 5.7±6.7 mm Hg (range, 2-12 mm Hg) without significant pulmonary regurgitation. There was no adverse event associated with the valve. Conclusions-A feasibility study of the Pulsta valve for native right ventricular outflow tract lesions was completed successfully with planned Pulsta valve implantation and demonstrated good short-term effectiveness without serious adverse events.

Original languageEnglish
Article numbere006494
JournalCirculation: Cardiovascular Interventions
Volume11
Issue number6
DOIs
StatePublished - 1 Jun 2018

Fingerprint

Pulmonary Valve
Pulmonary Valve Insufficiency
Swine
Feasibility Studies
Stroke Volume
Republic of Korea
Tetralogy of Fallot
Pulmonary Artery
Heart Ventricles
Stents
nitinol
Clinical Trials
Pressure

Keywords

  • feasibility study
  • heart diseases
  • nitinol
  • pulmonary valve
  • stent

Cite this

@article{70cd411fc3194c4cb7929a1e36d7123c,
title = "Successful Feasibility Human Trial of a New Self-Expandable Percutaneous Pulmonary Valve (Pulsta Valve) Implantation Using Knitted Nitinol Wire Backbone and Trileaflet-Gal-Free Porcine Pericardial Valve in the Native Right Ventricular Outflow Tract",
abstract = "Background-Self-expandable percutaneous pulmonary valve implantation (PPVI) for native right ventricular outflow tract lesions is still in the clinical trial phase. The aim of this study is to present the result of feasibility study of a novel self-expandable knitted nitinol wire stent mounted with a treated trileaflet-Gal-free porcine pericardial valve for PPVI. Methods and Results-A feasibility study using Pulsta valve (TaeWoong Medical Co, Gyeonggi-do, South Korea) was designed for patients with severe pulmonary regurgitation in the native right ventricular outflow tract, and 6-month follow-up outcomes were reviewed. Ten tetralogy of Fallot patients were enrolled. Before PPVI, severe pulmonary regurgitation (mean pulmonary regurgitation fraction, 45.5{\%}±7.2{\%}; range, 34.9{\%}-56{\%}) and enlarged right ventricular volume (mean indexed right ventricular end-diastolic volume, 176.7±14.3 mL/m2; range, 158.9-205.9 mL/m2) were present. The median age at PPVI was 21.7±6.5 years (range, 13-36 years). Five patients were successfully implanted with 28 mm and the other 5 with 26 mm valves loaded on the 18F delivery cable. No significant periprocedural complications were noted in any patient. At the 6-month follow-up, indexed right ventricular end-diastolic volume was dramatically decreased to 126.3±20.3 mL/m2 (range, 99-164.2 mL/m2), and the mean value of peak instantaneous pressure gradient between the right ventricle and the pulmonary artery decreased from 6.8±3.5 mm Hg (range, 2-12 mm Hg) before PPVI to 5.7±6.7 mm Hg (range, 2-12 mm Hg) without significant pulmonary regurgitation. There was no adverse event associated with the valve. Conclusions-A feasibility study of the Pulsta valve for native right ventricular outflow tract lesions was completed successfully with planned Pulsta valve implantation and demonstrated good short-term effectiveness without serious adverse events.",
keywords = "feasibility study, heart diseases, nitinol, pulmonary valve, stent",
author = "Kim, {Gi Beom} and Song, {Mi Kyoung} and Eunjung Bae and Park, {Eun Ah} and Whal Lee and Lim, {Hong Gook} and Kim, {Yong Jin}",
year = "2018",
month = "6",
day = "1",
doi = "10.1161/CIRCINTERVENTIONS.118.006494",
language = "English",
volume = "11",
journal = "Circulation: Cardiovascular Interventions",
issn = "1941-7640",
publisher = "Lippincott Williams and Wilkins Ltd.",
number = "6",

}

TY - JOUR

T1 - Successful Feasibility Human Trial of a New Self-Expandable Percutaneous Pulmonary Valve (Pulsta Valve) Implantation Using Knitted Nitinol Wire Backbone and Trileaflet-Gal-Free Porcine Pericardial Valve in the Native Right Ventricular Outflow Tract

AU - Kim, Gi Beom

AU - Song, Mi Kyoung

AU - Bae, Eunjung

AU - Park, Eun Ah

AU - Lee, Whal

AU - Lim, Hong Gook

AU - Kim, Yong Jin

PY - 2018/6/1

Y1 - 2018/6/1

N2 - Background-Self-expandable percutaneous pulmonary valve implantation (PPVI) for native right ventricular outflow tract lesions is still in the clinical trial phase. The aim of this study is to present the result of feasibility study of a novel self-expandable knitted nitinol wire stent mounted with a treated trileaflet-Gal-free porcine pericardial valve for PPVI. Methods and Results-A feasibility study using Pulsta valve (TaeWoong Medical Co, Gyeonggi-do, South Korea) was designed for patients with severe pulmonary regurgitation in the native right ventricular outflow tract, and 6-month follow-up outcomes were reviewed. Ten tetralogy of Fallot patients were enrolled. Before PPVI, severe pulmonary regurgitation (mean pulmonary regurgitation fraction, 45.5%±7.2%; range, 34.9%-56%) and enlarged right ventricular volume (mean indexed right ventricular end-diastolic volume, 176.7±14.3 mL/m2; range, 158.9-205.9 mL/m2) were present. The median age at PPVI was 21.7±6.5 years (range, 13-36 years). Five patients were successfully implanted with 28 mm and the other 5 with 26 mm valves loaded on the 18F delivery cable. No significant periprocedural complications were noted in any patient. At the 6-month follow-up, indexed right ventricular end-diastolic volume was dramatically decreased to 126.3±20.3 mL/m2 (range, 99-164.2 mL/m2), and the mean value of peak instantaneous pressure gradient between the right ventricle and the pulmonary artery decreased from 6.8±3.5 mm Hg (range, 2-12 mm Hg) before PPVI to 5.7±6.7 mm Hg (range, 2-12 mm Hg) without significant pulmonary regurgitation. There was no adverse event associated with the valve. Conclusions-A feasibility study of the Pulsta valve for native right ventricular outflow tract lesions was completed successfully with planned Pulsta valve implantation and demonstrated good short-term effectiveness without serious adverse events.

AB - Background-Self-expandable percutaneous pulmonary valve implantation (PPVI) for native right ventricular outflow tract lesions is still in the clinical trial phase. The aim of this study is to present the result of feasibility study of a novel self-expandable knitted nitinol wire stent mounted with a treated trileaflet-Gal-free porcine pericardial valve for PPVI. Methods and Results-A feasibility study using Pulsta valve (TaeWoong Medical Co, Gyeonggi-do, South Korea) was designed for patients with severe pulmonary regurgitation in the native right ventricular outflow tract, and 6-month follow-up outcomes were reviewed. Ten tetralogy of Fallot patients were enrolled. Before PPVI, severe pulmonary regurgitation (mean pulmonary regurgitation fraction, 45.5%±7.2%; range, 34.9%-56%) and enlarged right ventricular volume (mean indexed right ventricular end-diastolic volume, 176.7±14.3 mL/m2; range, 158.9-205.9 mL/m2) were present. The median age at PPVI was 21.7±6.5 years (range, 13-36 years). Five patients were successfully implanted with 28 mm and the other 5 with 26 mm valves loaded on the 18F delivery cable. No significant periprocedural complications were noted in any patient. At the 6-month follow-up, indexed right ventricular end-diastolic volume was dramatically decreased to 126.3±20.3 mL/m2 (range, 99-164.2 mL/m2), and the mean value of peak instantaneous pressure gradient between the right ventricle and the pulmonary artery decreased from 6.8±3.5 mm Hg (range, 2-12 mm Hg) before PPVI to 5.7±6.7 mm Hg (range, 2-12 mm Hg) without significant pulmonary regurgitation. There was no adverse event associated with the valve. Conclusions-A feasibility study of the Pulsta valve for native right ventricular outflow tract lesions was completed successfully with planned Pulsta valve implantation and demonstrated good short-term effectiveness without serious adverse events.

KW - feasibility study

KW - heart diseases

KW - nitinol

KW - pulmonary valve

KW - stent

UR - http://www.scopus.com/inward/record.url?scp=85053923702&partnerID=8YFLogxK

U2 - 10.1161/CIRCINTERVENTIONS.118.006494

DO - 10.1161/CIRCINTERVENTIONS.118.006494

M3 - Article

VL - 11

JO - Circulation: Cardiovascular Interventions

JF - Circulation: Cardiovascular Interventions

SN - 1941-7640

IS - 6

M1 - e006494