Durable-polymer versus biodegradable-polymer drug-eluting stents in acute coronary syndromes: three-year outcomes of the HOST REDUCE POLYTECH RCT Trial

the HOST REDUCE POLYTECH RCT Trial investigators

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

BACKGROUND: Long-term follow-up is essential to evaluate the impact of polymer degradation in drug-eluting stents (DES). AIMS: We aimed to compare durable-polymer DES (DP-DES) and biodegradable-polymer DES (BP-DES) during a 3-year follow-up to evaluate the entire period of polymer resolution (before, during, and after degradation). METHODS: The HOST REDUCE POLYTECH RCT Trial was a randomised clinical trial enrolling patients with acute coronary syndrome (ACS) and comparing the efficacy and safety of DP-DES and BP-DES. The primary outcome was a patient-oriented composite outcome (POCO), and the key secondary outcome was a device-oriented composite outcome (DOCO). RESULTS: A total of 3,413 ACS patients were randomised to either the DP-DES (1,713 patients) or BP-DES (1,700 patients) group. During the 3-year follow-up, the risk of the POCO was similar between the DP-DES and BP-DES groups (14.8% vs 15.4%, hazard ratio [HR] 0.96, 95% confidence interval [CI]: 0.80-1.14; p=0.613). However, the risk of the DOCO was lower in the DP-DES group (6.0% vs 8.0%, HR 0.73, 95% CI: 0.57-0.95; p=0.020). In a landmark analysis, the lower risk of the DOCO for the DP-DES group was evident during the transition from the early to the late period after percutaneous coronary intervention (PCI) (from 8 to 16 months post-PCI; 1.8% vs 3.3%, HR 0.54, 95% CI: 0.34-0.84; p=0.007), which was mainly driven by a risk reduction of target lesion revascularisation. CONCLUSIONS: In ACS patients, DP-DES showed similar results to BP-DES regarding the POCO up to 3 years. For the DOCO, DP-DES were superior to BP-DES; this was due to the higher event rate during the period of polymer degradation.

Original languageEnglish
Pages (from-to)e750-e759
JournalEuroIntervention
Volume20
Issue number12
DOIs
StatePublished - Jun 2024

Bibliographical note

Publisher Copyright:
© 2024 Europa Group. All rights reserved.

Keywords

  • ACS/NSTE-ACS
  • coronary artery disease
  • drug-eluting stent

Fingerprint

Dive into the research topics of 'Durable-polymer versus biodegradable-polymer drug-eluting stents in acute coronary syndromes: three-year outcomes of the HOST REDUCE POLYTECH RCT Trial'. Together they form a unique fingerprint.

Cite this