Clinical Efficacy and Safety of an Automatic Closed-Suction System in Mechanically Ventilated Patients with Pneumonia: A Multicenter, Prospective, Randomized, Non-Inferiority, Investigator-Initiated Trial

Dong Hyun Joo, Hyo Chan Park, Joon Han Kim, Seo Hee Yang, Tae Hun Kim, Hyung Jun Kim, Myung Jin Song, Sung Yoon Lim, Sung A. Kim, Hee Won Bae, Yoon Hae Ahn, Si Mong Yoon, Jimyung Park, Hong Yeul Lee, Jinwoo Lee, Sang Min Lee, Jung Chan Lee, Young Jae Cho

Research output: Contribution to journalArticlepeer-review

Abstract

Endotracheal suctioning is an essential but labor-intensive procedure, with the risk of serious complications. A brand new automatic closed-suction device was developed to alleviate the workload of healthcare providers and minimize those complications. We evaluated the clinical efficacy and safety of the automatic suction system in mechanically ventilated patients with pneumonia. In this multicenter, randomized, non-inferiority, investigator-initiated trial, mechanically ventilated patients with pneumonia were randomized to the automatic device (intervention) or conventional manual suctioning (control). The primary efficacy outcome was the change in the modified clinical pulmonary infection score (CPIS) in 3 days. Secondary outcomes were the frequency of additional suctioning and the amount of secretion. Safety outcomes included adverse events or complications. A total of 54 participants, less than the pre-determined number of 102, were enrolled. There was no significant difference in the change in the CPIS over 72 h (−0.13 ± 1.58 in the intervention group, −0.58 ± 1.18 in the control group, p = 0.866), but the non-inferiority margin was not satisfied. There were no significant differences in the secondary outcomes and safety outcomes, with a tendency for more patients with improved tracheal mucosal injury in the intervention group. The novel automatic closed-suction system showed comparable efficacy and safety compared with conventional manual suctioning in mechanically ventilated patients with pneumonia.

Original languageEnglish
Article number1068
JournalDiagnostics
Volume14
Issue number11
DOIs
StatePublished - Jun 2024

Bibliographical note

Publisher Copyright:
© 2024 by the authors.

Keywords

  • critical care
  • endotracheal suctioning
  • intensive care unit
  • mucosal secretions

Fingerprint

Dive into the research topics of 'Clinical Efficacy and Safety of an Automatic Closed-Suction System in Mechanically Ventilated Patients with Pneumonia: A Multicenter, Prospective, Randomized, Non-Inferiority, Investigator-Initiated Trial'. Together they form a unique fingerprint.